| Samples | Whole blood or hemolysis sample |
|---|---|
| Meas. items | HbA1c (Stable HbA1c), HbF (HbS and HbC can be detected in Variant mode) |
| Reagents *1 | ELUENT 80A, ELUENT 80B, ELUENT 80CV and HEMOLYSIS WASHING SOLUTION 80H |
| Column *2 | COLUMN UNIT 80 |
| Meas. principle | Reversed-phase cation exchange chromatography |
| Detection method | Dual-wavelength colorimetry (wavelengths measured: 420nm/500nm) |
| Resolution | 0.1% Ratio, 1 mmol/mol |
| Meas. range *3 | HbA1c : 3-20%, 9-195 mmol/mol HbF : 0-100% |
| Guaranteed meas. range *4 | HbA1c : 4-16%, 20-151 mmol/mol HbF : 0.3-5% |
| Processing speed | Variant mode : 90 seconds per sample Fast mode : 48 seconds per sample |
| Sample consumption | Approx.14μL (whole blood) |
| Sample container | Blood collection tube (12.3 to 15mm diameter) × (75 to 100mm length) Sample cup (500μL) |
| Sampling method | Cap-Piercing |
| Rack type | ARKRAY racks |
| Number of measurement samples | One-way transportation (factory set) : Maximum 50 samples Loop transportation : Maximum 100 samples |
| Column temperature | Approx. 40℃ |
| Display | Color graphic LCD with backlight |
| Memory capacity | 900 test results (including calibration results) |
| Built-in Printer | Thermal printer, 58mm thermal paper |
| External output | Serial 1 port (Can be optionally used as an Ethernet port.) |
| Meas. conditions | Temperature : 10-30℃ Humidity : 20-80% relative humidity (non-condensing) |
| Required sample volume | Blood collection tube : Minimum 10mm away from the bottom Sample cup : 400μL or more |
| Warm-up time | Maximum 30 min. |
| Dimensions | 530(W) × 530(D) × 530(H)mm (Not including the hemolysis washing solution bottle) |
| Weight | Analyzer: Approx. 39kg, Sampler: Approx. 4kg |
| Power supply | AC100V-240V ±10% 50/60Hz |
| Power consumption | Max. 300 VA |
![]()