Automatic Glycohemoglobin Analyzer
ADAMS A1c HA-8180V

The fastest measurement in the world
The HA-8180V incorporates unique HPLC column technology to make it the fastest analyzer of its kind in the world.

  • Variant Mode
    Time until first report is just 3.5 minutes. For subsequent samples, measurement takes just 90 seconds.

  • Fast Mode
    Time until first report is just 1.9 minutes. For subsequent samples, measurement takes just 48 seconds.

Easy-to-switch
Switching between Variant mode and Fast mode is performed via the operator panel on the analyzer, making things simpler for the user.
The same reagents and column are used for both Variant and Fast modes.

Highly-accurate measurements

data_out.jpg
The highly accurate measurements are the result of the HPLC measurement principle. The measurement result includes information of each peak and chromatogram. The result will be shown in both IFCC and NGSP units.

*In addition to measuring HbA1c and HbF, HA-8180V can detect variant Hb (HbS and HbC) in the Variant mode.


Easy-to-follow user interface:
Save time, avoid errors.

interface.jpg
A large-size color LCD displays actual result, shows analyzer status, comments to operation procedure or informs the user about the remaining reagent volume.


Sample-stirring function

sample_stirring.jpg
The automatic sample-stirring function before measurement prevents fluctuation in measurement results caused by blood cell sedimentation.


Easy-to-maintain

maintain.jpg
No special tools are required for daily maintenance. Exchanged parts are hand-tightened, minimizing the workload of the user. The pre-filter is already integrated with the column eliminating the hassle of exchanging it. Replacement of the filter is not required throughout the column life.


Cap-piercing method

cap_piercing.jpg
The HA-8180V uses a cap-piercing method which enables the setting of capped blood collection tubes in sample racks.
The accurate and direct aspiration of blood samples not only reduces workload but also helps to protect lab technicians from infection.


Specifications

Samples Whole blood or hemolysis sample
Meas. items HbA1c (Stable HbA1c),
HbF (HbS and HbC can be detected in Variant mode)
Reagents *1 ELUENT 80A, ELUENT 80B, ELUENT 80CV and HEMOLYSIS WASHING SOLUTION 80H
Column *2 COLUMN UNIT 80
Meas. principle Reversed-phase cation exchange chromatography
Detection method Dual-wavelength colorimetry
(wavelengths measured: 420nm/500nm)
Resolution 0.1% Ratio, 1 mmol/mol
Meas. range *3 HbA1c : 3-20%, 9-195 mmol/mol
HbF : 0-100%
Guaranteed meas. range *4 HbA1c : 4-16%, 20-151 mmol/mol
HbF : 0.3-5%
Processing speed Variant mode : 90 seconds per sample
Fast mode : 48 seconds per sample
Sample consumption Approx.14μL (whole blood)
Sample container Blood collection tube
(12.3 to 15mm diameter) × (75 to 100mm length)
Sample cup (500μL)
Sampling method Cap-Piercing
Rack type ARKRAY racks
Number of measurement samples One-way transportation (factory set) : Maximum 50 samples
Loop transportation : Maximum 100 samples
Column temperature Approx. 40℃
Display Color graphic LCD with backlight
Memory capacity 900 test results (including calibration results)
Built-in Printer Thermal printer, 58mm thermal paper
External output Serial 1 port
(Can be optionally used as an Ethernet port.)
Meas. conditions Temperature : 10-30℃
Humidity : 20-80% relative humidity (non-condensing)
Required sample volume Blood collection tube : Minimum 10mm away from the bottom
Sample cup : 400μL or more
Warm-up time Maximum 30 min.
Dimensions 530(W) × 530(D) × 530(H)mm
(Not including the hemolysis washing solution bottle)
Weight Analyzer: Approx. 39kg,
Sampler: Approx. 4kg
Power supply AC100V-240V ±10% 50/60Hz
Power consumption Max. 300 VA
*1, *2 Reagents and column are sold separately from HA-8180V analyzer.
*3 Error-free measurement ranges.
*4 Guaranteed ranges for obtaining results equivalent to those from HA-8160 VP mode measurement.

**Design and specifications may be changed without prior notice

Download for more information

HA-8180V